Wondcon 610A Infant Incubator
Wondcon 610A
Accuracy and durability,Mindry quality,competitive price!
PRODUCT DETAILS
Reliable
TEMPERATURE MONITORING AND PROTECTION ALARMS
EcoLa series incubator monitors and controls temperature on various points of heating system to ensure accuracy and safety. The alarms will be actived if any unmoral phenomenon is monitored.
Air Probe Failed | This alarm occurs when one of the three thermistors In the sensor module differ from the other by 0.8°C or both two thermistors have a measurement error exceeding acceptable limlts.thermlstors In the sensor module differ from the other by 0.8°C or both two thermistors have a measurement error exceeding acceptable limits. |
Heater Failed 1 | This alarm occurs when the heater Is open or shorted. |
Heater Failed 2 | This alarm occurs when the heater thermocouple wires are open or shorted. |
Humidity Heater Failed 1 | This alarm occurs when the humidity heater Is open or shorted. |
Humidity Heater Failed 2 | This alarm occurs when the heater thermocouple wires are open or shorted. |
EcoLa series incubator monitors the condition of air, humidity, oxygen sensors, air flow probes and sensor box to ensure all the data collected are true and accurate.
Sensor Disconnect | This alarm occurs when communication with the sensor module falls. |
Sensor Module Failure 3 | This alarm occurs when the controller detects a sensor module fan Is not rotating. |
Sensor Module Failure 6 | This alarm occurs when the controller detects a temperature multiplexer calibration follure. |
Sensor Out of Position | This alarm occurs when the sensor module Is not In the hood. |
Air How Probe Failed | This ahrm occurs If the air flow probe connection Is open or short circuited. |
Oxygen Cell Difference | This alarm occurs when the oxygen cell readings differ by more than 3%. As a result, the oxygen flow into the system Is interrupted. |
In order to keep proper running and provide the best user experience, EcoLa series incubator monitors the working condition of each aspect.
Stuck Key | This alarm occurs when a stuck key Is detected. |
Motor Failed | This alarm Is activated when the fan motor speed falls outside |
Low Air Flow | This alarm occurs when an air circulation failure is |
Add Water | No enough water In water reservoir. |
Water Reservoir Not Engaged | This alarm occurs when the water reservoir not In |
AMBIENT TEMPERATURE MONITORING
With ambient temperature probe EcoLa series incubator is able to compare the gap between ambient and setting temperature and changing heating power output accordingly in order to provide more stable inside hood air temperature.
HEATING SYSTEM MONITORING
EcoLa monitors not only hood air temperature but also radiator and circuit air temperature. By collecting each parts' temperature information, EcoLa's temperature control system is much more accurate and safer than others.
PID FUZZY ALGORITHM
With the help of latest PID technology, EcoLa minimizes the temperature variability to unprecedented 0.1 °C and temperature overshoot is also reduced dramatically.
TEMPERATURE UNIFORMITY
Comfortable
•EcoLa provides newborn a supper quiet environment. With the help of aerodynamics EcoLa's hood sound is reduced to 45dB
•With great detail design, EcoLa is also devoted to provide medical personnel pleasure operation experience.
Specification
Physical Attributes | Height 126.1 cm to 146.1cm | Width 99.7cm | Depth 57.7cm Weigh 80 kg |
Hood Specifications | Tubing access ports 10 | Access door size 18 x 13cm | Mattress to hood height 4.8cm |
Soft bed mattress size 73.6 x 38.6 x1.8cm | Mattress tilt ±12° | Drawer depth 36cm | |
Drawer size 28.0 x 24.7cm 28.0 x 11.3cm(x2) | |||
Temperature Control Modes | Temperature control modes Skin and airtemperature control mode | Air mode control temperature range 20°C-37°C | Air mode control override temperature range 37°C-39°C |
Air mode control accuracy 37°C-39°C | Skin mode control temperature range <±1.0°C | Skin mode control override temperature range 34°C-37°C | |
Skin mode control override temperature range | Skin temperature accuracy ±0.3% | Dual-skin temperature monitoring Yes | |
Servo Humidity Option | Humidity control range 30%-95%RH | Humidity control operating time without refilling 24 hours | Humidity control reservoir capacity 1500ml |
Humidity display accuracy ±5% RH | rol accuracy Humidity control accuracy ±5% RH |
Performance | Air flow velocity across mattress <10cm/sec | Temperature rise time at 22 °C (72 °F) ambient <20min(From 22°C,50% RH,to 35°C) | Temperature variability <0.5 °C |
Temperature overshoot < 0.5 °C maximum | Temperature uniformity with a level mattress <0.8 °C | Operating noise level in hood ≦45dBa | |
Carbon Dioxide (CO2) Level <0.5% | |||
Servo Oxygen Option | Oxygen display accuracy of full scale ±2% | Oxygen control accuracy (100% calibration) ±3% | Oxygencontrol accuracy (21% calibration) ±5% |
Oxygen control range 21 % to 65% | Oxygen display resolution 1% | ||
Scale Option | Weight range 300 g to 8 kg | Weight display resolution 1g | Weight accuracy ±10 g |
Operating Environment | Temperature 20° to 30°C | Humidity 10 to 95%RH | Air Velocity Up to 0.3 m/sec |
Storage/Shipping Information | Temperature -25° C to 60° C | Humidity 0 to 95% RH |