Wondcon WME980G ETCO2 monitor
Wondcon WME980G
compact,stable and reliability portable Patient monitor.
PRODUCT DETAILS
Features
1.5 inch TFT, monitoring the EtC02/C02, Respiration rate
Wondcon's EtC02 technology of infrared ray gas analysi
Temperature, Pressure compensation and Zero readjustment system, all of them insure the accuracy and stability
Balance gas compensation makes minimum influence against other gases,
Lung pressure and temperature compensation (BTPS compensation) is selectable
Very short warmup time, in15 seconds it can measure with complete accuratcy
Easy maintenance and no need to calibrate
KPA, mmHg or % are selectable.
Solid design challenges hush conditions and specially designed airway adapter is suitable for intubated adult or infant monitoring.
Audio-optical alarm for exceeding limits and Apnea
24 hours trend diagram keeps stilt available after power-off
Data output via USB port
Optioanl data base software administration analysis and printing data
Optional bluetooth module and bluetooth printer, at any time to print wave shape and data
Large Capacity lithium battery for 10 hours UPS
Applied situations
Pre-hospital first aid, hospital transit, emergency department, operating theater, iCU, NICU, sleep monitoring etc.
Specification
Power | |
AC input: | 100V-250V, 50HZ/60 Hz AC |
power Consumption: | 5W |
Operation | |
Working temperature: | 5 -40 "C |
Humidity: | 30 -75% ( non-condensing) |
Atmospheric pressure: | 86-106kPa |
Battery Data | Lithium battery, rechargeable |
Battery capacity: | 3.6V,1200mAh; 10 hours (approximately) |
Charging time: | 4 hours from empty to full |
Physical Data | |
Dimension: | 55mm(W)x44mm(H) x44mm(D) |
Weight: | 80 grams |
Capnograph Measuring Data | |
EtC02/C02 | |
Range: | 0-150mmHg < 0-20kPa,or 0-20% <V/V> |
Accuracy: | ±2mmHg when 0-40mmHg, ±5%(reading) when > 41-70mmHg ±8%(reading) when > 71-100mm Hg ±10%(reading) when > 101-150mmHg |
EtC02 Update: | every breath, |
Trend length: | 24hours |
Respiration rate | |
Range: | 3 -150 t/min |
Accuracy: | the bigger value of ±l%(reading) or ±ltime$/min |
Trend length: | 24 hours |
Monitoring and measuring ventilation status and circulation functions, locating cannula, alerting lung hypoventilation/ hyperventilation, cooperating to guide the application of anaesthetic machine/ ventilator, in time to discover contingency, CPR tracking, non-invasive evaluation of PAC02, distinguishing and classifying urgently injured patients etc