Wondcon WME980D Patient monitor
Wondcon WME980D
compact,stable and reliability portable Patient monitor. It is used for hospital,clinic,etc
PRODUCT DETAILS
Features
7” high resolution color TFT display
Arrhythmia analysis and ST segment analysis
Up to 12000 groups data record storage
60-hour ECG waveform storage and recall
1000-hour data trends for all parameters with graphic and list view
2000 groups of event data storage
2000 groups of arrhythmia events data with corresponding HR, TEMP, CO2, SpO2, and RESP value
Protection against defibrillator interference
Adult / Infant measurement modes
Visual and audible alarms
Product Accessories
NIBP | |
Technique | Oscillometric |
Cuff pressure range | 0 mmHg ~ 300mmHg |
Pressure precision | ±3mmHg |
Cuff inflation time | <10 seconds (typical adult cuff) |
Overpressure protection | Adult/Pediatric 300mmHg Neonate 140mmHg |
BP measuring range | Systolic pressure Adult 40mmH~255mmHg Pediatric 40mmHg~200mmHg |
Mean arterial pressure | Adult 20mmHg~215mmHg Pediatric 20mmHg~165mmHg |
Diastolic pressure | Adult 10mmHg-195mmHg Pediatric 10mmHg~150 mmHg |
BP measuring accuracy | Maximal mean difference ±5 mmHg |
Maximal standard deviation | 8 mmHg |
Measurement mode | Manual, Auto, STAT |
TEMP | |
Measuring range | 25.0°C~45.0°C |
Measuring accuracy | ±0.2°C |
Responding time | W150s |
Quick temp(optional) | <20s(skin), <2s(ear) |
SpO2 | |
Transducer | Dual-wave length LED |
SpO2 measuring range | 35%~100% |
Low perfusion capability | 0.4%~5% |
SpO2 measuring accuracy | Arms less than 3% for SpO2 range from 70% to 100% |
Low perfusion performance | 0.4% |
PR measuring range | 30bpm~240bpm |
PR measuring accuracy | ±2bpm or ±2%, whichever is greater |
ECG | |
Input dynamic range | ±0.4mVp~±5mVp |
HR measuring range | 20bpm~300bpm |
HR measuring accuracy | ±1 % or ±2bpm, whichever is greater |
HR alarm delay time | W10s |
Sensitivity selection | *1/2, *1, *2, tolerance £5% |
Sweeping speed | 12.5mm/s, 25mm/s, 50mm/s, tolerance M10% |
ECG noise level | ^30|iVp-p |
ECG input loop current | WO.IpA |
Input impedance | 25MQ |
Common-mode rejection ratio | N89dB |
Time constant | 20.3s (Monitoring mode) |
Frequency response | 0.5Hz~40Hz (Monitoring mode) |
RESP | |
RR measuring range | 0qjm~120rpm |
RR measuring accuracy | ±5% or ±2 rpm, whichever is greater |
Others | |
Power supply | 100~240VAC, 50/60Hz |
Rechargeable Li-ion battery | 14.4V 2200mAh |
Display | 7 inches TFT color LCD with 800x480 |
Alarming | Audible & visual alarm |
Communication | Serial/ Ethernet |
Standard Configuration | |
ECG, RESP, SpO2, NIBP, TEMP, PR | |
Option | |
Nellcor SpO2, Trolley Wall Mount, Central monitoring system, External printer |